Specifications

Monoxor
Plus Manual Calibration and Maintenance
0019-9370 Rev 0 47
Step CO Span Procedure
3 Attach a 500 ppm CO cylinder to the regulator and connect calibration kit
components as shown below. Apply 500 ppm carbon monoxide in an air
balance calibration gas.
4 Wait until the Measured reading stabilizes and then press ENTER. The
message “Good Calibration” should briefly appear.
If the sensor’s output is low, but still usable, then the message “Good
Calibration WARNING Low Sensor” will appear. The sensor will now be
marked as being Low in the Warm-up screen.
If the sensor’s output is too low to be usable, then the message “Bad
Calibration Sensor End of Life, Entry Not Saved” will appear.
5 Close the regulator and remove the CO cylinder.
5.7. T-Ref Sensor Calibration
The T-Ref sensor is located inside the instrument. Calibration is done at the
factory and should not need to be done in the field.