Specifications

Monoxor
Plus Manual Calibration and Maintenance
0019-9370 Rev 0 45
Step Temperature Calibration Procedure
value that exactly equals the setting of the simulator.
Alternatively: Submerge probe tip into a container of boiling water with a
thermometer, wait several minutes, and then use the arrow buttons to enter
an Applied value that exactly equals the thermometer reading.
NOTE: The calibration range is from 175 to 625° F (80 to
329° C). An attempt to calibrate outside this range will cause
the message “Applied Value High (or Low) to appear at the
bottom of the screen.
7 Wait until the Measured reading stabilizes, and then press ENTER to calibrate
the TS-Span Measured value to that of the “Applied” value, after which the
message “Good Calibration” should briefly appear followed by the
CALIBRATION menu being re-displayed.
5.6. CO Sensor Calibration
5.6.1. Materials Required
Calibration kit, P/N 0024-7059
Gas cylinder: 500 ppm CO in air, P/N 0024-0492
5.6.2. CO Manual Zero Procedure
The CO zeroing process is done automatically during warm-up or manually
using the manual zero feature (see page 25). To perform a manual zero, follow
the steps below. If your instrument is configured for CO auto zero, then skip
this CO manual zero procedure and go to the CO Sensor Span procedure that
follows.
Step Manual CO Zero Procedure
1 If not already done, turn ON the analyzer and display the Main Menu screen.
2 Use the UP () and DOWN () arrow buttons to select the SETUP menu and
press ENTER.
3 From the Setup Menu, use the UP () and DOWN () arrow buttons to
select the CO Zero Setting parameter then press ENTER.