Section 5. Calibration and Maintenance
5.1. Serviceability
The instrument operator is able to easily replace the following components
without the use of tools:
• probe assembly
• probe filters
• batteries
• B-Smart
CO sensor
• printer paper.
Additionally, a technician, with the use of factory-provided instructions, can:
• perform basic diagnostics
• confirm proper operation
before putting the unit back into service. Field calibration is also possible with
the proper equipment. Refer to the calibration section for more information.
5.2. Cleaning the Probe
The probe tube and gas sample hose will become dirty under normal use.
NOT
E: The filter element should prevent soot from reaching the
analyzer’s internal components. If the probe is not kept clean, it
could become clogged and restrict the flow of gas into the
analyzer, resulting in incorrect test readings.
NOTE:
An analyzer that tests natural gas furnaces normally
requires less frequent cleaning than an analyzer used for testing
coal- or oil-fired furnaces.