Step Example Procedure for Taking Differential Temperature Measurements
8 At this point, you may press SAVE again to save the differential
temperature value to memory so that it can be retrieved and/or printed
later.
9 Carefully remove the probe assembly and allow it to cool.
NOT
E:
A negative differential value will be displayed if T1<T2.
4.7. Timed Ambient CO Testing
This procedure takes approximately 15 minutes to complete and provides a
minute-by-minute snapshot of CO readings, as well as a “Max CO” value that
represents the highest CO reading measured during the entire 15-minute test.
Results can be saved to memory and/or printed. Use the optional CO
Appliance Kit (
P/N 0024-8555) and the following procedure to perform an
ambient CO measurement.
Step Example Procedure for Taking Ambient CO Measurements
1 Attach optional appliance kit probe (P/N 0024-8555).
2
Turn on the Monoxor
Plus in a fresh air environment and wait for
initialization to complete.
3
Verify successful initialization (no errors).
4
Check battery status (see page 12). If battery life is questionable,
replace the batteries, as the Ambient CO test takes approximately
15 minutes to complete.
5
Move instrument to target location to be tested.
6
Press F2 to display the Main Menu.
7
Use the down arrow to highlight Ambient CO Test and press the
ENTER button.
8
Follow the on-screen instructions to initiate the test.
Refer to page 16 for details on navigating the ambient CO test
screens, viewing results, saving results to memory, and printing
results.