Specifications
Operation Monoxor
Plus Manual
32 0019-9370 Rev 0
NOT
E: Use the left or right arrow buttons from the Run or Hold
screen to view the CO trending graph.
4.4. CO Trending Graph
A trending screen is accessible from the RUN or HOLD screen by using the right
of left arrow keys. The CO Trending Screen shows dynamic CO levels graphed
over a programmable time period. User-selectable time periods are:
• 30 seconds
• 1 minute
• 3 minutes
• 5 minutes
• 15 minutes.
The g
raph continues to sample CO gas while in RUN mode, showing the current
PPM reading above the dynamic graph. Press the RUN/HOLD button (HOLD
mode) to freeze the current graphical snap-shot. In HOLD mode, a pointer
appears on the graph. This pointer can be moved along the graph using the left
and right arrow keys. As you move the pointer left and right, the value of the
CO at that point in the sampling is displayed at the top of the screen.
4.5. Ending a Test
WARNING: Burn Hazard. Do not touch the probe after removing
it from a flue. Allow the probe to cool before handling (about 5
minutes).
After taking a gas sample, remove the probe and take the analyzer to an area
containing fresh air. Allow the pump to run until the CO reading drops to near
zero.
netzerotools.com