Specifications
Configuration Fyrite
InTech
Manual
34 0024-9486 Rev 3
Setup Menu Function
threshold value (in 100 ppm increments) before
returning to the Main Menu.
During combustion analysis, if the Protect CO feature is
enabled and the CO reading equals or exceeds the CO
Limit setting, the analyzer shuts off the pump and
prompts the operator for a course of action.
• Continue (and risk sensor damage)
• Purge
Continue Option
• Pump starts.
:
• Future High CO warnings are suppressed until CO
drops below 100 ppm.
• After CO drops below 100 ppm, the analyzer
begins enforcing the current Protect CO threshold
limit again.
Purge Option
• Remove probe tip from flue and press ENTER.
:
• Pump starts.
• Purge begins.
• Measured CO is displayed.
• Purge ends when CO level drops below 50 ppm.
The Combustion Air Temp option allows you to select a
source for the combustion air temperature (T-Air) used
by the analyzer for calculations. There are 3 choices.
1: T-Air value is read
dynamically from the
temperature sensor inside
the analyzer.
2: Temporarily use probe
(T-STK) to get T-AIR value.
When ready, press ENTER
to save displayed reading.
Value is saved and used as
fixed T-AIR value. After
save, the probe
(thermocouple) is used for
taking T-STK readings
again.