Step  T-Air Calibration Procedure 
NOTE: The calibration  range is from 32 to 41° F (0 to 5° C). 
An attempt to calibrate outside this range will cause the 
message “Applied Value High
” (or Low) to appear at the 
bottom of the screen. 
5  Wait until the measured reading stabilizes, and then press ENTER to calibrate 
the TA-Zero Measured value to that of the applied value, after which the 
message “Good Calibration” should briefly appear followed by the CALIBRATE 
TA-SPAN screen. 
6 
Set thermocouple  simulator to 212° F (100° C), and then use the UP (▲), 
DOWN (▼), LEFT (◄), and RIGHT (►) arrow buttons  to enter an applied 
value that exactly equals the setting of the simulator. 
Alternatively:  Submerge probe  tip into a container of boiling water  with a 
thermometer, wait several minutes, and then use the UP (▲), DOWN (▼), 
LEFT (◄), and RIGHT (►) arrow buttons  to enter an applied value that 
exactly equals the thermometer reading. 
NOTE: The calibration range is from 194 to 230° F (90 to 
110° C). An attempt to calibrate outside this range will cause 
the message “Bad Calibration Wrong CAL Entry” to appear in 
the following step. 
7  Wait until the measured reading stabilizes, and then press ENTER to calibrate 
the TA-Span Measured value to that of the applied value, after which the 
message “Good Calibration
” should briefly appear followed by the 
CALIBRATION MENU screen being re-displayed. 