Instruction manual
Fyrite
 INSIGHT
 Plus Manual  Calibration and Maintenance 
0024-9487 Rev 2  91 
5.7.2. T-Air Calibration Procedure 
Step  T-Air Calibration Procedure 
1 
Plug the simulator into the T-AIR  connector located at the bottom of the 
analyzer. 
Alternatively:  Plug the probe’s thermocouple  into the T-AIR  connector 
located at the bottom of the instrument.  
IMPORTANT: DO NOT attach the probe’s gas hose  to the 
analyzer’s GAS port, otherwise water will be drawn into the 
analyzer! 
2  If not already done, turn ON the analyzer and 
display the CALIBRATION MENU. Note that 
this requires password  validation (see page 
55). 
3  Use the UP (▲) and DOWN (▼) buttons to highlight T-Air, and then press 
ENTER to display the CALIBRATE TA-ZERO screen. 
NOTE: “Measured” is the current temperature reading, while 
“Applied” is a known temperature that will be applied for 
calibration purposes. 
4 
Set thermocouple simulator to 32° F (0° C), 
and then use the UP (▲), DOWN (▼), LEFT 
(◄), and RIGHT (►) arrow buttons to enter an 
applied value that exactly equals the setting of 
the simulator. 
Alternatively: Submerge probe tip into an ice-water bath with a 
thermometer, wait several minutes, and then use the UP (▲), DOWN (▼), 
LEFT (◄), and RIGHT (►) arrow buttons to enter an applied value that 
exactly equals the thermometer reading. 