Step  T-STACK Calibration Procedure 
6  Set thermocouple simulator to 572° F (300° C), and then use the UP (), 
DOWN (), LEFT (), and RIGHT () arrow buttons to enter an Applied 
value that exactly equals the setting of the simulator. 
Alternatively:  Submerge probe  tip into a container of boiling water  with a 
thermometer, wait several minutes, and then use the arrow buttons to enter 
an Applied value that exactly equals the thermometer reading. 
NOTE: The calibration  range is from 175 to 625° F (79 to 
329° C). An attempt to calibrate outside this range will cause 
the message “Applied Value High” (or Low) to appear at the 
bottom of the screen. 
7  Wait until the Measured reading stabilizes, and then press ENTER to calibrate 
the TS-Span Measured value to that of the “Applied” value, after which the 
message “Good Calibration” should briefly appear followed by the 
CALIBRATION screen being re-displayed. 
5.7. T-Air Calibration 
This procedure first zeros and then spans T-AIR temperature to known 
temperature values.  
The use of an electronic thermocouple  simulator is the preferred method of 
producing the desired calibration temperatures. Alternatively, containers of ice 
water and boiling water can be used. 
5.7.1. Materials Required 
•
Thermocouple Simulator (K-type) 
Range:   0 to 600° F  (-18 to 316° F) 
Accuracy:  ± 0.5° F  (± 0.3° C) 
•
(Alternatively) Ice Water, Boiling Water, Thermometer 