Step  Pressure Sensor Calibration Procedure 
7  Wait until the Measured reading stabilizes, and then press ENTER to calibrate 
the pressure sensor's Measured value to that of the Applied value. The 
message, "Good Calibration" should briefly appear, followed by the 
CALIBRATION menu. 
8  Remove calibration equipment. 
5.6. T-Stack Calibration 
This procedure first zeroes and then spans stack temperature  to known 
temperature values. 
The use of an electronic thermocouple  simulator is the preferred method of 
producing the desired calibration  temperatures. Alternatively, ice and boiling 
water baths can be used. 
5.6.1. Materials Required 
•
Thermocouple simulator (K-type) 
Range:   0 to 600° F (-18 to 316° F) 
Accuracy:  ± 0.5° F (± 0.3° C) 
•
(Alternatively) ice water, boiling water, thermometer 
5.6.2. T-Stack Calibration Procedure 
Step  T-STACK Calibration Procedure 
1  Plug the simulator into the T-STACK connector located at the bottom of the 
analyzer. 
Alternatively: Plug the probe's thermocouple into the T-STACK connector 
located at the bottom of the analyzer.  
IMPORTANT: DO NOT attach the probe's gas hose  to the 
analyzer's GAS port; otherwise water will be drawn into the 
analyzer! 