Instruction manual
Fyrite
 INSIGHT
 Plus Manual  Calibration and Maintenance 
0024-9487 Rev 2  87 
Step  Pressure Sensor Calibration Procedure 
1  Assemble the pressure sensor calibration equipment as shown above, but DO 
NOT connect the analyzer to the calibration equipment at this time. 
2  If not already done, power ON the analyzer and display the CALIBRATION 
menu. Note that this requires password validation (see page 55). 
3  Use the UP () and DOWN () arrow buttons to select PRESSURE and then 
press ENTER to display the CALIBRATE PRESSURE screen. 
"Measured" is the pressure value currently being detected by the pressure 
sensor, while "Applied” is a known value of pressure that will be applied for 
calibration purposes. 
4  With both the -ΔP and +ΔP ports open to the atmosphere, observe that the 
current measured pressure reading should be 0.00 ± 0.01 inwc. If necessary, 
zero the pressure sensor  (Menu  →  Pressure  →  Zero) then repeat steps 2 
through 4). 
5  Connect the hose from the manometer to the +ΔP port and apply a negative 
pressure  to this port by adjusting the bellows  for a manometer reading of 
-4.00 (negative 4.00). Other units are shown below. 
Units  Name 
Nominal Calibration 
Point 
inwc  inches water column  -4.00 inwc 
mB  millibars  -10.00 mB 
hPa  hecto Pascals  -10.00 hPa 
Pa  Pascals  -1000 Pa 
mm H
2
O  millimeters of water  -101.6 mm H
2
O 
6  Use the UP (), DOWN (), LEFT (), and RIGHT () arrow buttons to enter 
an “Applied” value that exactly equals the manometer reading. 
NOTE: The calibration range is from -6 to -2 inwc (-15 to -5 mB). 
An attempt to calibrate outside this range will cause the 
message "Applied Value High" (or Low) to appear at the bottom 
of the screen. 