Instruction manual
Fyrite
 INSIGHT
 Plus Manual  Calibration and Maintenance 
0024-9487 Rev 2  85 
5.4.5. B-SMART
 CO Sensor Replacement 
Step 
B-SMART
 CO Sensor Replacement 
1  Enter the CALIBRATION MENU. Note that this requires password  validation 
(see page 55). 
2  Use the UP () and DOWN () arrow buttons to select B-Smart. Press 
ENTER to display the B-Smart code screen. 
3  Use the UP () and DOWN () arrow buttons  to enter the 10-digit 
alphanumeric code supplied with the pre-calibrated B-SMART
 sensor. Use 
the  LEFT  ()  and RIGHT () arrow  buttons  to move the cursor across the 
screen. Press ENTER. 
NOTE: If the correct code was entered, the analyzer accepts it 
and returns to the CALIBRATION MENU.  If an incorrect code 
was entered, the screen will display "Invalid Code." Check to 
make sure the correct code has been entered. If problem 
persists, contact your nearest Bacharach Service Provider. 
NOTE: B-SMART
 codes can be entered through the Fyrite
User Software (FUS). 
NOTE: Installing a B-SMART
 sensor forces the instrument to perform a 
zero function (either manual or automatic). 
NOTE: Bacharach offers a convenient Exchange 
Program  (where available) 
that allows the 
customer to regularly receive pre-calibrated 
replacement sensors  that include a code that can 
be entered into the analyzer for a quick convenient 
setup. Contact Bacharach customer service for 
more details about this program. 