Instruction manual
Fyrite
 INSIGHT
 Plus Manual  Calibration and Maintenance 
0024-9487 Rev 2  79 
Section 5.  Calibration and Maintenance 
5.1. Serviceability 
The instrument operator is able to easily replace the following components 
without the use of tools: 
•  probe assembly 
•  probe filters 
•  batteries 
•  printer paper. 
Additionally, a technician, with the use of readily available hand tools and 
factory-provided instructions, can: 
•  perform basic diagnostics 
•  replace sensors 
•  confirm proper operation 
before putting the unit back into service. Field calibration is also possible with 
the proper equipment. Refer to the calibration section starting on page 86 for 
more information. 
5.2. Cleaning the Probe 
The probe tube and gas sample hose will become dirty under normal use. 
NOTE: The water trap’s filter  element should prevent soot  from 
reaching the analyzer’s internal components. If the probe  is not 
kept clean, it could become clogged and restrict the flow of gas 
into the analyzer, resulting in incorrect combustion  test readings 
and calculations. 
NOTE: 
An analyzer that tests natural gas furnaces normally 
requires less frequent cleaning  than an analyzer used for testing 
coal- or oil-fired furnaces. 










