Instruction manual
Fyrite
 INSIGHT
 Plus Manual  Overview 
0024-9487 Rev 2  13 
NOTE: The  North American (NA) configuration of the Fyrite
INSIGHT
 Plus computes and displays the calculations as long as the 
measured oxygen is not above 16% O
2
 and the stack temperature is 
not above 650° C (1202° F). The Siegert configuration of the Fyrite
INSIGHT
 Plus computes and displays the calculations as long as the 
measured oxygen is not above 18.8% O
2
 and the stack temperature 
is not above 650° C (1202° F). 
Calculation  Calculation Range 
Reso-
lution 
Configuration 
NA  Siegert 
Efficiency (HHV)  0.1 to 100 %  0.1%  X  X 
ETA (LHV)  0 to 115%  0.1%    X 
Excess Air  1 to 250 %  1%  X   
Stack Loss  0.1 to 100 %  0.1 %    X 
Lambda  1 to 9.55  0.01    X 
CO
2
 (dry basis)  0.1 to a fuel-dependent max in %  0.1 %  X  X 
CO Ref to O
2
  0 to 9999 ppm  1 ppm  X  X 
CO/CO
2
 Ratio  0.0001 to fuel-dependent max  0.0001    X 
∇ ∇ ∇ 