Instruction manual
Overview  Fyrite
 INSIGHT
 Plus Manual 
12  0024-9487 Rev 2 
Specification  Description 
IrDA Port  Protocol:  IrDA-SIR    Data Bits: 8 
Baud Rate:  9600      Stop Bits: 1 
Parity:  None 
Memory  100 locations for storing test results 
Power Supply 
Options 
Batteries 
(4 AA) 
Type:  Disposable Alkaline (Included) 
Duration:  15 hours min, continuous max draw 
Type:  Disposable Lithium 
Duration:  20 hours, continuous max draw 
Type:  Rechargeable 
Duration:  8 hours, continuous max draw 
USB Cable 
(A to Mini B) 
Source:  PC 
Source:  AC source (via Wall Adapter) 
Measure-
ment 
Range  Resolution  Accuracy 
Response 
Time (T
90
) 
O
2
 and 
LL O
2
0 to 20.9 %  0.1% O
2
± 0.3% O
2
< 20 sec 
CO w/ NO
x
filter 
0 to 
4000 ppm 
1 ppm 
± 10 ppm  (0 to 200 ppm)  
± 5%   (201 to 4000 ppm) 
< 40 sec 
Ambient 
Temp 
-20° to 316° C 
(-4° to 600° F) 
0.1° C 
(0.1° F) 
± 1° C   (0 to 100° C) 
< 70 sec 
Stack Temp 
-20° to 650° C 
(-4° to 1202° F) 
1° C 
(1° F) 
± 2° C   (0° to 124° C) 
± 3° C   (125° to 249° C) 
± 4° C   (250° to 400° C) 
< 50 sec 
Differential 
Temp 
± 600° C 
(± 1112° F) 
0.1° C 
(0.1° F) 
N/A  N/A 
Pressure / 
Differential 
Pressure 
± 100 mB 
(± 40 inwc) 
0.01 mB 
(0.01 inwc) 
± 0.03 mB   ( -1 to 1 mB) 
± 3%   (-40 to -1 mB) 
± 3%   (1 to 40 mB) 
N/A 