Section 6.  Troubleshooting 
6.1. Error and Warning Messages 
Message  Description 
T-STK 
Disconnected 
The probe  thermocouple  is not connected to the analyzers 
T-Stack connector. Plug the probe thermocouple plug into the 
T-Stack connector at the bottom of the instrument. 
Check Sensor 
O
2
O
2
 sensor output is low, but still usable. Sensor may need to be 
replaced in the near future. The arrow on the O
2
  Sensor Life 
screen is in the “replace” segment. Refer to page 56. 
Replace Sensor 
O
2
O
2
  sensor  output is low and should be replaced. The arrow on 
the O
2
 Sensor Life screen is beyond the bar graph (typically 2 
years for standard O
2
  sensors and 3 years for long-life O
2
sensors). Refer to page 56. 
Bad Sensor O
2
  O
2
 sensor output is too low and is not usable. 
Low Sensor CO  CO sensor output was low but still usable. Sensor may need to be 
replaced in the near future. 
Low Battery  Battery voltage is low. Replace the batteries. 
Applied Value 
High/Low 
An attempt was made to calibrate  a sensor outside its range—
either above (High) or below (Low) the acceptable range. 
Warmup Sensor 
Error 
•  CO sensor was not zeroed at warmup because of high output. 
Run instrument on fresh air then restart instrument to re-zero 
sensor.  If the message persists, the CO sensor may need to be 
replaced. 
•  Stack or Air temperature sensors are measuring temperature 
outside the range of -4° to 212° F at warmup. Make sure that 
the Stack and Air thermocouples are sampling ambient room 
air within the temperature range at warmup. 
•  The  Fyrite
  INSIGHT
  Plus was turned on with the probe 
sampling flue gas. Move the probe to fresh air and restart the 
instrument. 
•  Messages will indicate which sensors are in error. 