Instruction manual
Calibration and Maintenance  Fyrite
 INSIGHT
 Plus Manual 
94  0024-9487 Rev 2 
5.8.3. CO Sensor Span Procedure 
Step  CO Span Procedure 
1  From the Calibration Menu, use the UP () and DOWN () arrow buttons to 
highlight CO, and then press ENTER to display the CALIBRATE CO screen. 
Note that this requires password validation (see page 55). 
“Measured” is the current CO  reading, while “Applied” is a known CO level 
that will be applied for calibration purposes. 
2  Use the UP (), DOWN (), LEFT (), and RIGHT () arrow buttons to enter 
an Applied value that exactly equals the concentration stamped on the CO 
cylinder. 
NOTE: Bacharach recommends using a 500 ppm calibration gas, 
however the calibration range is from 20 to 1,000 ppm. An 
attempt to calibrate outside this range will cause the message 
“Applied Value High” (or Low) to appear at the bottom of the 
screen. 
3  Attach a 500 ppm CO  cylinder to the regulator  and connect calibration  kit 
components as shown below. Apply 500 ppm carbon monoxide in an air 
balance calibration gas. 
4  Wait until the Measured reading stabilizes and then press ENTER. The 
message “Good Calibration” should briefly appear. 
If the sensor’s output is low, but still usable, then the message “Good 
Calibration WARNING Low Sensor”  will appear. The sensor will now be 
marked as being Low in the Warm up screen. 
If the sensor’s output is too low to be usable, then the message “Bad 
Calibration Sensor End of Life, Entry Not Saved” will appear. 
5  Close the regulator and remove the CO cylinder. 