Specifications

Instruction 0011-9026 Page 7
10. This step completes CO
2
or O
2
FYRITE gas sample analysis. A few points to re-
member when reading the FYRITE:
(a) FYRITE accuracy is within ±1/2% CO
2
or O
2
compared to actual value.
(b) Always handle FYRITE by ns to ensure body heat is not absorbed by uid.
(c) A delay in reading of 5 or 10 seconds may decrease accuracy of reading
slightly but longer delays may cause substantial error.
(d) The FYRITE is calibrated to indicate on a “dry” basis for ue gas samples,
which are normally fully saturated with moisture. Failure to arti cially
moisten wool packing in Filter Saturator Tube when actual sample is not
fully saturated will cause FYRITE to read slightly low.
Therefore, make certain wool packing in Filter Saturator Tube is suf ciently
moistened with water for accurate results on non- ue gases. Avoid excessive
moisture which can be drawn into the Aspirator Bulb and forced into the
FYRITE during sampling.
CAUTION
Never depress plunger valve to vent FYRITE in the inverted
position. This will cause uid to spill, which is corrosive and
contains poisonous elements. In the event of a spill, read In-
structions in Section 2.2 FYRITE uid handling precautions.
3.1 Determining CO
2
and O
2
FYRITE Fluid Strength
FYRITE Fluid strength can be conveniently checked immediately after the rst
sample reading. Without venting, repeat the absorbing operations (Steps 7 through
9) by inverting FYRITE again and positioning upright until all uid drains to the bot-
tom reservoir. Observe scale reading. An increase of more than 1/2% CO
2
or O
2
in the
second reading as compared to the rst indicates a need for uid replacement.
3.2 Operating Precautions to Ensure Maximum Accuracy
Locate top of uid column (refer to Fig. 11). The surface at the top of the uid column
(meniscus) in the small, center bore will be dish-shaped as shown in Figure 11. Either
the high or low point of this dish- shaped surface may be used to locate top of uid col-
umn providing the same point is always used both for setting scale zero and reading
percent CO
2
or O
2
. Obviously, using high point for one operation and low point for the
other will cause an error.
Best practice is to use high point of this uid surface just at the small center bore
wall. In setting scale zero or reading percent CO
2
or 0
2
, hold FYRITE vertically and
level with eyes and sight across scale to top of uid column.