7. Absorb sample gas into
FYRITE by inverting
until uid drains into top
reservoir (Fig. 7). Then
turn upright (Fig. 8) to
drain uid into bottom
reservoir. Repeat this
step once.
8. Hold FYRITE at 45°
angle (Fig. 9) momentari-
ly to allow uid droplets
to drain into the bottom
reservoir.
9. With FYRITE held
upright (Fig. 10), per-
mit uid in column to
stabilize a few seconds,
then immediately read %
carbon dioxide on scale at
the point corresponding to
top of the uid column.
CO
2
FYRITE OPERATION O
2
FYRITE OPERATION
(RED FLUID) (BLUE FLUID)
7. Absorb sample gas into
FYRITE by inverting
until uid drains into top
reservoir (Fig. 7). Then
turn upright (Fig. 8) to
drain uid into bottom
reservoir. Repeat this
step three (3) more times
(four complete inversions
total).
8. Hold FYRITE at 45°
angle (Fig. 9) momentari-
ly to allow uid droplets
to drain into the bottom
reservoir.
9. With FYRITE held
upright (Fig. 10), per-
mit uid in column to
stabilize a few seconds,
then immediately read
% oxygen on scale at the
point corresponding to top
of the uid column.
Figure 7
Figure 8
Figure 9
Figure 10