Operating instructions
FLORITE
®
800
INSTRUCTION 16-9013
Installation/Operation/Maintenance
Rev. 3 – March 2004
Bacharach, Inc.
621 Hunt Valley Circle, New Kensington, PA 15068
Phone: 724-334-5000 • Fax: 724-334-5001 • Web: www.bacharach-inc.com
Printed in U.S.A.
® Registered Trademarks of
Bacharach, Inc.
Introduction
The Florite 800 is a battery powered, hand held instrument with a digital LCD
display capable of measuring the velocity, volume flowrate and temperature
of air by means of a thermal probe. The unit allows the airstream area to be
entered for volume measurement. Measurements may be displayed in either
Metric or Imperial units as selected from the keyboard. After selection, this
will be ‘remembered’ by the instrument until reselected from the keyboard. The
attached thermal anemometer probe has a tip diameter of 1/4" (8mm) and is,
therefore, suitable for measurements at grilles, slot diffusers and in small ducts
where large access holes are not acceptable. The velocity sensor in the probe
is compensated for variations in airstream temperature by means of a second
thermistor which also senses, and displays, the actual airstream temperature in
either metric or imperial units consistent with the units selected for velocity.
The instrument is supplied with batteries. Four 1.5 volt ‘AA’ size cells are re-
quired which may be either disposable or rechargeable types. Either Standard
or Alkaline disposable cells may be used but the Standard type will exhibit a
relatively short life and are, therefore, not recommended. Low battery condition
is indicated by ‘LOBAT’ appearing on the instrument display. The cells must be
replaced as soon as possible to maintain optimum accuracy. The battery cells are
accessible by sliding the cover (at th e rear of the instrument case) in the direc-
tion of the arrow. A tape is provided in the battery compartment for easy removal
of the battery cells. Ensure that the battery cells are replaced with the correct
polarity as indicated in each cell compartment. Rotate the cell slightly forwards
and backwards to check that the cell is properly seated in the cell compartment.








