User Guide

Fyrite
InTech
Manual Calibration and Maintenance
0024-9486 Rev 4 55
Step
CO Sensor Replacement Procedure
the diagram below, then twisting the cap
clockwise approximately 40 to the
"closed" position (2 o’clock).
7
Reattach tubing.
8
Calibrate the CO sensor using either the
standard calibration procedure or the B-
SMART
procedure
5.4.5. B-SMART
CO Sensor Replacement
Step
B-SMART
CO Sensor Replacement
1
Enter the CALIBRATION MENU. Note that this requires password validation
(see page 35).
2
Use the UP () and DOWN () arrow buttons to select B-Smart. Press ENTER
to display the B-Smart code screen.
3
Use the UP () and DOWN () arrow buttons to enter the 10-digit
alphanumeric code supplied with the pre-calibrated B-SMART
sensor. Use
the LEFT () and RIGHT () arrow buttons to move the cursor across the
screen. Press ENTER.
NOTE: If the correct code was entered, the analyzer accepts it
and returns to the CALIBRATION MENU. If an incorrect code was
entered, the screen will display "Invalid Code." Check to make
sure the correct code has been entered. If problem persists,
contact your nearest Bacharach Service Provider.
NOTE: B-SMART
codes can be entered through the Fyrite
User
Software (FUS).
NOTE: Installing a B-SMART
sensor forces the instrument to perform a
zero function (either manual or automatic).