User Guide
Fyrite
InTech
Manual Overview
0024-9486 Rev 4 11
1.10. Specifications
Specification
Description
Temperature
Storage: -20 to 50 C ( -4 to 122 F)
0 to 20 C (32 to 68 F ) optimal
Operation: -5 to 45 C (23 to 113 F)
Reference: 20 ± 2 C (68 ± 4 F)
Humidity
Storage: 15 to 90% RH, non-condensing
Operation: 15 to 95% RH, non-condensing
Reference: 45 ± 10% RH, non-condensing
Pressure
1 atmosphere ± 10%
Weight
16 ounces (454 g) with batteries
Dimensions
8.0” x 3.6” x 2.3” (20.3 cm x 9.1 cm x 5.8 cm) (H x W x D)
Warm-up Time
Minimum = 30 seconds; Maximum = 60 seconds
Gas Sample Flow
Rate
300 to 700 cc/min
Sensors
O
2
Electrochemical (P/N: 0024-0788)
CO Electrochemical (P/N: 0024-7265)
Temp (Stack) K-Type thermocouple
Product Approvals
and Regulatory
Compliance
EN50270: (CE Mark) EMC tested in accordance with
European Directive 2004/108/EC .
EN50379: Standard for portable electrical apparatus
designed to measure combustion flue gas
parameters of heating appliances (Siegert only)
Parts 1 and 3.
RoHS Compliance
Case Construction
High impact ABS plastic with co-molded rubber.
Protective rubber boot with molded-in magnets.
Display
Monochrome with backlight
USB Connector
Mini B (USB 2.0)
Memory
10 locations for storing test results
IrDA Port
Protocol: IrDA-SIR Data Bits: 8 Stop Bits: 1
Baud Rate: 9600 Parity: None
Power Supply
Options
Batteries
(4 AA)
Type: Disposable Alkaline (Included)
Duration: 15 hours min, continuous max draw
Type: Disposable Lithium
Duration: 20 hours, continuous max draw
Type: Rechargeable
Duration: 8 hours, continuous max draw