Step CO Sensor Replacement Procedure
below, then twisting the cap
clockwise approximately 40° to the
"closed" position (2 o’clock).
7 Reattach tubing.
8 Calibrate the CO sensor using either
the standard calibration procedure or
the B-SMART
procedure
5.4.5. B-SMART
CO Sensor Replacement
Step
B-SMART
CO Sensor Replacement
1 Enter the CALIBRATION MENU. Note that this requires password validation
(see page 32).
2 Use the UP () and DOWN () arrow buttons to select B-Smart. Press
ENTER to display the B-Smart code screen.
3 Use the UP () and DOWN () arrow buttons to enter the 10-digit
alphanumeric code supplied with the pre-calibrated B-SMART
sensor. Use
the LEFT () and RIGHT () arrow buttons to move the cursor across the
screen. Press ENTER.
NOTE: If the correct code was entered, the analyzer accepts it
and returns to the CALIBRATION MENU. If an incorrect code
was entered, the screen will display "Invalid Code." Check to
make sure the correct code has been entered. If problem
persists, contact your nearest Bacharach Service Provider.
NOTE: B-SMART
codes can be entered through the Fyrite
User Software (FUS).
NOTE: Installing a B-SMART
sensor forces the instrument to perform a
zero function (either manual or automatic).