Instruction manual

98
Transmitter Cond Ind 7100e/2(X)H
CSA Control Drawing
Conductivity Measuring Loop
Terminals
1, 2, 3, 4, 5, 6
Uo, Vsc Io, Isc Po Co, Ca Lo, La
IIC (GRP A, B) 9V 111mA 139mW F 1.5mH
IIB (GRP C) 9V 111mA 139mW 12µ F 6mH
IIC (GRP D) 9V 111mA 139mW 32µ F 12mH
Temperature Measuring Loop
Terminals
7, 8
Uo,Vsc Io,Isc Po Co,Ca Lo,La
IIC (GRP A, B) 5V 3.5mA 5mW 100µF 1H
IIB (GRP C) 5V 3.5mA 5mW 300µF 1H
IIC (GRP D) 5V 3.5mA 5mW 800µF 1H
All Combined Outputs
Terminals
1, 2, 3, 4, 5, 6, 7, 8
Uo, Vsc Io, Isc Po Co, Ca Lo, La
IIC (GRP A, B) 9V 114.5mA 144mW F 1.5mH
IIB (GRP C) 9V 114.5mA 144mW 12µF 6mH
IIC (GRP D) 9V 114.5mA 144m W 32µF 12mH
70098_E_CI7100_2XH.QXD 27.07.2005 9:28 Uhr Seite 98