Instruction manual
83
Service functions
Current source Loop current specifiable 3.8 ... 22.00 mA
Device self-test Automatic memory test (RAM, FLASH, EEPROM)
Display test Display of all segments
Last Error Display of last error occurred
Sensor monitor Display of direct, uncorrected sensor signal
(resistance/temperature)
Passcodes Modifiable according to FDA 21 CFR Part 11
“Electronic Signatures”
Data retention Parameters and calibration data > 10 years (EEPROM)
EMC EN 61326
Emitted interference: Class B (residential area)
Class A
Immunity to interference: Industry
Explosion protection
Cond Ind 7100e/2XH: ATEX: TÜV 99 ATEX 1434
II 2 (1) G EEx ib [ia] IIC T6
FM: FMRC J.I. 300580
IS/I/1/ABCD/T4
NI/I/2/ABCD/T4
CSA: 1662790
CI I, Div 1, Gr ABC & D T4; Ex ib [ia] IIC T4
CI I, Div 2, Gr ABC & D, T4; Ex nAL[L] IIC T4
Cond Ind 7100e/2H: FM: FM 300580
NI/I/2/ABCD/T4
Nominal operating conditions
Ambient temperature –20 ... +55 °C
Transport/Storage temp –20 ... +70 °C
Supply voltage 14... 30 V