Instruction manual
81
Sensor standardization
Operating modes • Entry of cell factor with simultaneous display
of conductivity and temperature
• Input of conductivity of calibration solution
with simultaneous display of cell factor and
temperature
• Product calibration
• Zero point adjustment
• Temperature probe adjustment
Adm. cell factor 00.100 ... 19.999
Adm. transfer ratio 001.00 ... 199.99
Adm. zero point deviation ±0.5 mS/cm
Sensor monitoring Monitoring of primary coil and its lines
Sensocheck for short circuit and of secondary
and its lines for open circuit
Sensoface Provides information on the sensor condition
(zero point, Sensocheck)
Sensor monitor Direct display of measured values from sensor
for validation (resistance / temperature)
Temperature input
*)
Pt100 / Pt1000 / NTC 30 kΩ / NTC 100 kΩ
2-wire connection, adjustable
Ranges Pt100/Pt1000 -20 .. +200 °C
(-4 ... +392 °F)
NTC 30 kΩ -20 ... +150 °C
(-4 ... +302 °F)
NTC 100 kΩ -20 ... +130 °C
(-4 ... +266 °F)
Resolution 0.1 °C / 1 °F
Measurement error
1,2,3)
0.5 K (<1 K for Pt100; <1 K for NTC >100 °C)
Temperature compensation *
(Reference temp 25 °C) (OFF) none
(Lin) Linear characteristic 00.00 ... 19.99 %/K
(NLF) Natural waters to EN 27888
70098_E_CI7100_2XH.QXD 27.07.2005 9:28 Uhr Seite 81