Instruction manual
76
Transmitter Cond Ind 7100e/2(X)
Sensocheck
Continuously monitors the primary coil and its lines for short
circuits and the secondary coil and its lines for open circuits.
Critical values make the Sensoface “sad” and the correspon-
ding icon flashes:
The Sensocheck message is also output as error message Err 33
or Err 34. The red LED is lighted, the output current is set to
22 mA (when configured correspondingly). Sensocheck can be
switched off during configuration (then Sensoface is also
disabled). Exception: After a calibration a Smiley is always
displayed for confirmation.
Sensoface
(Sensocheck must have been activated during configuration.)
The smiley in the display (Sensoface) alerts to sensor problems
(defective sensor, defective cable). The conditions for a friendly,
neutral, or sad Sensoface are summarized in the following
chart. Additional icons refer to the error cause.
Note:
The worsening of a Sensoface criterion leads to the devaluation
of the Sensoface indicator (Smiley becomes “sad”).
To reset the Sensoface indicator, the defect must be remedied
and the transmitter be calibrated.
70098_E_CI7100_2XH.QXD 27.07.2005 9:28 Uhr Seite 76