Instruction manual
75
Sensocheck:
Primary coil
Sensocheck:
Secondary coil
Red LED
Out 1 (22 mA)
Current output
Current below 3.8 mA
Error
Icon
(flashes)
Problem
Possible causes
ERR 11
x
x
Current output
Current above 20.5 mA
ERR 12
x
x
Current output
Current span too small / too large
ERR 13
x
x
ERR 33
x
x
Temperature outside TC tables
(TC, Conc, SAL)
independent of
Sensoface
Sensoface
active
(see Pg 77)
Display of last error message
(Error Info)
Press conf while in measuring mode and confirm
code 0000*. The last error message is displayed for
approx. 20 sec.
After that the message will be deleted (immediate
return to measurement at pressing enter).
Display of last error message
Entry/
Display
Action / Remarks
ERR 34
0000*
* Factory setting
70098_E_CI7100_2XH.QXD 27.07.2005 9:28 Uhr Seite 75