Instruction manual

Measured
value
flashes
74
Transmitter Cond Ind 7100e/2(X)
Red LED
Out 1 (22 mA)
Error messages (Error Codes)
Sensor
• Measurement range violation
• SAL > 45 ‰
• Sensor connection or cable defective
• Wrong cell factor
Error
Display
Problem
Possible causes
ERR 01
Measured
value
flashes
x
x
Unsuitable sensor
Conductance range > 3000 mS
ERR 02
x
x
System error
Configuration or calibration data
defective. Completely reconfigure
and recalibrate the device.
Memory error in device program
ERR 98
“Conf”
flashes
x
x
Factory settings
EEPROM or RAM defective
This error message only occurs in
the case of a total defect. The
transmitter must be repaired and
recalibrated at the factory.
ERR 99
“FAIL”
flashes
x
x
Temperature probe
Open or short circuit
Temperature range exceeded
ERR 03
x
x