Instruction manual
71
Display of the output current
Press enter while in measuring mode.
For 5 sec, the secondary display shows the output
current instead of the temperature.
Diagnostics functions
Entry/
Display
Remark
Display of calibration data (Cal Info)
Press cal while in measuring mode and enter passcode
0000*. The current cell factor is shown in the main
display and the zero point underneath. After 20 sec
the transmitter returns to measuring mode (immediate
return at pressing enter).
Display of last error message (Error Info)
Press conf while in measuring mode and enter
passcode 0000*. The last error message is displayed
for approx. 20 sec.
After that the message will be deleted.
(immediate return to measurement at pressing enter).
0000*
2222*
0000*
Sensor monitor for validation
of sensor and complete measured-value processing.
Loop a defined sensing resistor (e.g. R = 100 Ω)
through the sensor as shown in the figure.
Press the conf key and enter code 2222*. The sensor
monitor displays the directly measured resistance and
the temperature. If there is a significant difference
between resistor value and display, the sensor and its
transmission behavior should be checked.
Press enter to return to measurement.
Note: The transmitter does not automatically go to
Hold mode.
* Factory setting