Instruction manual

70
Transmitter Cond Ind 7100e/2(X)
Display
Remark
Measurement
In the measuring mode the main display shows the
configured process variable (conductivity, concentration,
resistivity, or salinity) and the lower display the tempera-
ture. During calibration you can return to measuring
mode by pressing the cal key, during configuration by
pressing conf + enter (waiting time for measured-
value stabilization approx. 20 sec).
Display
Action Remark
Measure the temperature of the
process medium using an exter-
nal thermometer. Enter measured
temperature value:
Select with
key, edit number
with
key, proceed with
enter.
End adjustment with enter.
HOLD will be deactivated after
20 sec.
Temperature probe adjustment
Activate calibration
(Press cal,
Enter passcode 1015*.)
Select with
key, edit number
with
key, proceed with
enter.
Ready for calibration
Wrong settings
change the measure-
ment properties! If
an invalid passcode is
entered, the trans-
mitter returns to
measuring mode.
Default:
Current value of
secondary display.
Transmitter is in Hold
mode. Display for
approx. 3 sec
* Factory setting