Instruction manual
68
Transmitter Cond Ind 7100e/2(X)
Display
Action
Remark
When there has not
been an entry for
6 sec, the lower
display alternately
shows the conductiv-
ity and temperature
value.
Zero calibration with calibration solution
Calibration solution with low conductivity
Press cal key,
enter code 1001*
Select with key, edit number
with key, proceed with enter.
Immerse sensor in calibration
solution.
Modify the value until the lower
display shows the conductivity
value of the calibration solution.
Press enter to confirm calibra-
tion.
Transmitter is in Hold
mode.
If an invalid code is
entered, the trans-
mitter returns to
measuring mode.
Ready for calibration
Dismount and clean sensor
Display (3 sec)
The cell factor and zero point
are displayed.
Press enter to confirm the cali-
bration data.
* Factory setting
70098_E_CI7100_2XH.QXD 27.07.2005 9:28 Uhr Seite 68