Instruction manual

66
Transmitter Cond Ind 7100e/2(X)
Display
Action
Remark
When there has not
been an entry for
6 sec, the lower
display alternately
shows the conductiv-
ity and temperature
value.
Zero calibration in air
Press cal key,
enter code 1001*
Select with
key, edit number
with
key,
proceed with enter
Modify the zero point until zero
is displayed as conductivity value
in the lower display.
Select with
key, edit number
with
key.
If required, change the sign of
the zero point.
Press enter to confirm the zero
point.
Transmitter is in Hold
mode.
If an invalid code is
entered, the trans-
mitter returns to
measuring mode.
Ready for calibration
Dismount and clean sensor.
(Sensor must be dry!)
Display (3 sec)
* Factory setting