Instruction manual

Product calibration step 2:
When the sample value has
been determined, call up the
product calibration once more
(Press cal key. Enter passcode
1105*.
65
Display
Action
Remark
Measuring mode:
From the flashing CAL mode
indicator you see that product
calibration has not been termi-
nated.
Display
(approx. 3 sec)
Enter lab value. The new cell
factor is calculated.
While the sample
value is determined,
the transmitter
returns to measuring
mode.
The new cell factor and zero
point are displayed.
Confirm with enter.
The measured value is shown in
the main display alternately
with “Hold”; “enter” flashes.
End with enter.
Safety prompt.After
end of calibration,
the outputs remain
in Hold mode for
approx. 20 sec.
New calibration:
Press cal.
* Factory setting