Instruction manual
64
Transmitter Cond Ind 7100e/2(X)
The measurement process is only interrupted briefly.
During product calibration the sensor remains in the process.
Calibration is without TC correction!
Procedure: During sampling the currently measured value is stored in the
transmitter. The transmitter immediately returns to measuring mode. The
calibration mode indicator flashes and reminds you that calibration has not
been terminated. The sample is measured in the lab or directly on the site
using a portable meter. The measured sample value is then entered in the
transmitter. The new cell factor is calculated from these two values.
If the sample is invalid, you can take over the value stored during sampling.
In that case the old calibration values are stored. Afterwards, you can start
a new product calibration.
Product calibration
Calibration by sampling
Display
Action
Remark
Product calibration step 1:
Press cal key. Enter passcode
1105*.
(Press key to select position,
enter number using key,
confirm with enter)
Take sample and store value.
Proceed with enter
If an invalid passcode
is entered, trans-
mitter returns to
measuring mode.
The sample is meas-
ured in the lab or
directly on the site.
Display
(approx. 3 sec)
* Factory setting
70098_E_CI7100_2XH.QXD 27.07.2005 9:28 Uhr Seite 64