Instruction manual

60
Transmitter Cond Ind 7100e/2(X)
Display
Action
Remark
Enter cell factor:
Select with key,
edit number with key.
Conductivity and temperature
are alternately displayed during
the input (lower display).
Confirm entry with enter.
Calibration by input of cell factor
Press cal key, enter code 1100*
Select with
key, edit number
with
key,
proceed with enter
Ready for calibration
Dismount and clean sensor
If an invalid code is
entered, the trans-
mitter returns to
measuring mode.
The entered cell factor and zero
point are displayed.
Confirm with enter.
Display (3 sec)
Transmitter in Hold
mode, measured
value frozen.
Sensoface inactive.
Input of cell factor with simultaneous display of conductivity and
temperature (without temperature compensation)
* Factory setting
70098_E_CI7100_2XH.QXD 27.07.2005 9:27 Uhr Seite 60