Instruction manual

59
Information on calibration
Calibration can be performed by:
Entry of cell factor
Determining the cell factor with a known
calibration solution taking account of the temperature
Product calibration
Zero calibration in air or with calibration solution
Temperature probe adjustment
Note:
If measurements are taken in containers with
A < 110 mm, be sure to choose a container with
the same cross-section and the same material
(metal/plastic) for calibration.
Caution
All calibration procedures must be performed by trained
personnel. Incorrectly set parameters may go unnoticed, but
change the measuring properties.
When another sensor is used, its sensor data (cell factor,
transfer ratio, measuring frequency, temperature probe) must
be entered in the configuration menu before calibration.
Each time a new sensor is connected, the transmitter must
be calibrated.