Information on calibration
Calibration can be performed by:
• Entry of cell factor
• Determining the cell factor with a known
calibration solution taking account of the temperature
• Product calibration
• Zero calibration in air or with calibration solution
• Temperature probe adjustment
Note:
If measurements are taken in containers with
A < 110 mm, be sure to choose a container with
the same cross-section and the same material
(metal/plastic) for calibration.
Caution
• All calibration procedures must be performed by trained
personnel. Incorrectly set parameters may go unnoticed, but
change the measuring properties.
• When another sensor is used, its sensor data (cell factor,
transfer ratio, measuring frequency, temperature probe) must
be entered in the configuration menu before calibration.
• Each time a new sensor is connected, the transmitter must
be calibrated.