Instruction manual
53
OFF
(OFF
LIN
nLF)
Code
Display
Action
Choices
tc.
Only with linear temperature com-
pensation (LIN) selected:
Enter temperature coefficient*.
Select position with arrow key
and edit number with key.
Proceed with enter
02.00%/K
(XX.XX %/K)
Temp compensation selection
(not with Conc, Sal)
OFF: Temperature compensation
switched off. Select with key,
proceed with enter
LIN: Linear temperature compen-
sation with entry of temp coeffi-
cient and reference temperature.
nLF: Temperature compensation
for natural waters to EN 27888
°C
(°F)
Specify temperature unit
Select with
arrow key
Proceed with enter
* Reference temp 25 °C
70098_E_CI7100_2XH.QXD 27.07.2005 9:27 Uhr Seite 53