Instruction manual
36
Transmitter Cond Ind 7100e/2(X)
Overview of configuration steps
7250 IPR, Other
mS/cm, S/m, SAL, %
LIN / LOG
tc.UnIT
Select temperature unit
°C / °F
Select temperature compensation
(not for SAL)
OFF/LIN/NLF (natural waters)
Lin: Enter temperature coefficient
xx.xx %/K (02.00 %/K)
tc
Temperature compensation
Menu
Code
out1
Current output
o1.SnSR
Sensor selection
Select measured variable
Select solution
Codes: -01- to —10- see Pg 42
Characteristic linear / logarithmic
(not for SAL, Conc)
Selection / Default
ALrt
Alarm settings
AL.SnSO
Select Sensocheck
ON / OFF
Enter alarm delay
0000 ... 0600 SEC (0010 SEC)
LED in HOLD mode
ON / OFF
(Factory setting bold print)
o1.4mA
o1.20mA
o1.4mA
o1.UnIT
o1.CoNC
o1.CHAR
o1.20mA
tc.
tc.lin
AL.dLY
AL.LED
NaCl
-01-
Fix:
Time constant of output filter
xxxx SEC (0000 SEC)
22 mA signal for error messages
ON / OFF
Signal behavior during HOLD
Last / Fix
Enter fixed value
xxx.x mA (021.0 mA)
xxxx mS (000.0 mS)
Enter current end
LOG:
Enter current start
Enter current end
xxxx mS (000.0 mS)
in decades: 0.001 ... 1000 mS
(0.100 mS)
in decades: 0.001 ... 1000 mS
(100.0 mS)
LIN:
Enter current start
o1.FtME
o1.FAIL
o1.HoLD
o1.FIX
Codes -02- ... -10-
o1.CELL
Only with Other:
Entry of cell factor
Enter transfer ratio
Select temperature probe
2.175 (00.100...20.000)
120.00 (001.00...200.00)
Pt100/Pt1000/NTC100/NTC30
o1.SFC
o1.rTD
70098_E_CI7100_2XH.QXD 27.07.2005 9:27 Uhr Seite 36