Instruction manual

29
Safety functions
Sensocheck, Sensoface sensor monitoring
Sensocheck continuously monitors the sensor and lines.
The primary and its lines are monitored for short circuit and
the secondary and its lines are monitored for open circuit.
Sensocheck can be switched off (Configuration, Pg 55).
Sensoface provides information on the conductivity
sensor condition.
GainCheck device self-test
A display test is carried out, the software version is displayed
and the memory and measured value transfer are checked.
Start GainCheck device self test:
Automatic device self-test
The automatic device self-test checks the memory and meas-
ured-value transfer. It runs automatically in the background at
fixed intervals.
+
70098_E_CI7100_2XH.QXD 27.07.2005 9:27 Uhr Seite 29