Instruction manual
20
Transmitter Cond Ind 7100e/2(X)
Installation and connection
Information on installation
• Installation may only be carried out by trained experts in
accordance with this instruction manual and as per applicable
local and national codes.
• Be sure to observe the technical specifications and input ratings.
• Be sure not to notch the conductor when stripping the insulation.
• When commissioning, a complete configuration must be
carried out by the system administrator.
Connection to supply units
• Transmitter Cond Ind 7100e/2H: Before connecting this device
to a supply unit, make sure that its output voltage cannot exceed
30 V DC. Do not use alternating current or mains power supply!
• Transmitter Cond Ind 7100e/2XH: This device may only be
connected to an explosion-proof power supply unit (for input
ratings refer to annex of EC-Type-Examination Certificate).
Terminals: suitable for single wires / flexible leads up to 2.5 mm
2
(AWG 14)
The connections to the transmitter must be installed
in accordance with the National Electric Code (ANSI-
NFPA 70) Division 2 hazardous (classified) location
non-incendive wiring techniques.
FM Control Drawing: Refer to page
Division 2 wiring
94.
Warning!
Additional safety precautions have to be taken for applications in
hazardous locations to CSA!
(See Pg 97 et seq.)
70098_E_CI7100_2XH.QXD 27.07.2005 9:27 Uhr Seite 20