User Manual
1-66 
set or return parallel/series mode 
  :TYPE <MASTER/0,SLAVEA/1,SLAVEB/2,SLAVEC/3|?> 
set or return master/slave setting 
[SOURce] 
Source Subsystem 
  :CURRent <NRf | ? > 
set or return current level:0-40/0-24A 
    :PROTection <Boolean | ? > 
set or return over-current state 
      :LEVel <NRf | ? > 
set or return over-current level 
  :VOLTage <NRf | ? > 
set or return voltage level:0-36/0-60V 
    :PROTection <Boolean | ? > 
set or return over-voltage state 
      :LEVel <NRf | ? > 
set or return over-voltage level 
SYStem 
System Subsystem 
  :BEEP <Boolean | ? > 
set or return BEEP state 
  :E5V <Boolean | ? > 
enable or disable extra 5V output 
  :ERRor? 
return system error 
  :EXTernal<OFF/0,VOLT/1,RES/2 | ? > 
set or return external state 
  :GPIB 
    :ADDRess <NR1 | ? > 
set or return GPIB address (1-30) 
  :IP 
    :ADDRess <NR1.NR1.NR1.NR1 | ? > 
set or return IP address 
    :CONFig <STATic/0? > 
set or return IP config mode 
  :KEY 
    :LOCK <Boolean | ? 
set or return key lock state 