User Manual

1-56
5.4 Constant Voltage Protection (CV TO CC)
When this function is activated, the power supply will stay in CV mode. If load
changes force the system to transition from CV to CC (constant current) mode,
the system will enter the “CV TO CC Protect” state that will shut off the output
and display the CVC message on the screen. Press to reset the
protection and deactivate the buzzer.
5.5 Constant Current Protection (CC TO CV)
When this function is activated the power supply will stay in CC mode. If load
changes forces the transition from CC to CV (constant voltage) mode, the system
will enter the “CC TO CV Protect” state that will shut off output and display
CCV message on the screen. Press to reset the protection and deactivate
buzzer.
5.6 Over-temperature Protection (OTP)
When the instrument detects abnormally high temperature, the system will enter
the “Over Temperature Protect” mode that will shut off the output and display
the error message as shown in the following figure. Press to reset the
protection and deactivate buzzer.
=
36
.
.
= .
0
1
.
FF
L
!!!!
.
V = 36.00 V I = 40.00 A OFF
0.000 V 0.000 A
CVC