(or 0 - 24 A for XLN6024 or 0 - 18 A for XLN8018 or 0 - 14.4 A for XLN10014)
output current, the resolution of voltage and current will be different in response.
The screen will show similar to the following (accuracy is 10mV):
4.10 Timer of Current Flow
This function offers testing of the cutoff time of a breaker or a fuse. First, connect
the test object to output terminal as shown below.
Press
,
,
and then in the main screen to enter the current
flow counter screen.
Press VSET to set the voltage. Then press ISET to set the current (I), which is
the maximum current you want to test. Press ISET twice to set the breaker/fuse
current (Ib). After setting up the output voltage/current and pressing
to turn on the output, the system will start counting down the timer until the
breaker or fuse is open. The count starts from when output has reached the Ib
current. The resolution of the timer is 100us (0.1 ms) and the maximum
counting period is one hour. If the counter doesn’t work after pressing ,
error occurs herein and the screen will display an error message after two seconds.
= 1 . = .
:
.
FF= 1 .
:
V = 36.00 V I = 40.00 A OFF
0.00 V 0.00 A