After wiring is complete, configure one supply as the Master and the other three
as the Slave A, B, and C. After the power supply is configured as Master, it will
start searching for all Slaves that are connected to the Master. Therefore, in order
to setup correctly, user must set up the Slaves before the Master is set.
To set a XLN3640/XLN6024/XLN8018/XLN10014 to the slave mode, press
,
, and
in the main screen to enter the series/parallel
setting option. Then select the SERIES mode by turning the knob (SCPI
command is “PS:MODE SERIES”) and then press to confirm. It will
continue to the next line for the MASTER/SLAVE selection. Turn the knob to
select SLAVE A (SCPI command is “PS:TYPE SLAVEA”) for the instrument
and press to confirm the setting. Using the same procedure to set up
SLAVE B and C for the other two supplies.
To set the master unit, press
,
,
and
in the main screen to
enter the series/parallel setting option. Then use the knob to select the series
mode (SCPI command is “PS:MODE SERIES”) and press to confirm.
It will continue to the next line for the MASTER/SLAVE selection. Use knob
again to select MASTER (SCPI command is “PS:TYPE MASTER”) and then
press to confirm. After confirming the supply as the Master, it will start