User Manual

1-32
When Local sense is selected, the wire connection can be setup as follows
(shown in the figure above):positive sense (+S) is connected to the positive
lead (+) and negative sense (-S) is connected to the negative lead (-),
whereas the positive lead (+) of the DC output is connected to the positive
end (+) of the device and the negative lead (-) of the DC output is
connected to the negative end (-) of the device under test. When this
sensing mode is selected, the power wires from the DC output leads to the
device under test should be as short as possible.
(19) LAN (optional):
The ETHERNET interface connector
(20) GPIB (optional):
The GPIB interface connector
(21) AC power input:
The power receptacle is for a power source within 100 VAC - 240 VAC.
(22) Earth connection:
Used for earth ground connection.