User Manual
1-29
C. Please inspect the hardware if the current value is inconsistent with
the value in the tolerance error table below after the calibration
values of each function is changed to actual current value.
XLN3640
Function
Setting
Range of Transformation Value
Lo
0.1A
0 - 0.5 A
MIDL
13A
11 - 14 A
MIDH
26A
23 - 27 A
Hi
38A
34 - 39 A
XLN6024
Function
Setting
Range of Transformation Value
Lo
0.06A
0 - 0.5 A
MIDL
7.8A
6.6 - 9 A
MIDH
15.6A
14 - 17.2 A
Hi
22.8A
21 - 24.6 A
XLN8018
Function
Setting
Range of Transformation Value
Lo
0.045A
0 - 0.1 A
MIDL
5.9A
5.3 - 6.5 A
MIDH
12A
10.8 - 13.2 A
Hi
17A
15.3 - 18.7 A
XLN10014
Function
Setting
Range of Transformation Value
Lo
0.03625A
0 - 0.1 A