User Manual

1-23
8. CALIBRATION
Press
in the third page of Menu Setting to enter CALIBRATION
menu. Users must enter the password to access calibration mode.
8.1 Equipment Requirements
1. 5 ½ Digital Volt meter.
2. Shunt for current calibration (100 A/ 10 mΩ)
8.2 Calibration Procedure
VOLTAGE CALIBRATION
A. Connect power supply output terminal to DVM (as shown in
Figure 1 below). Turn on the supply. Once the unit enters the main
page, press MENU and select 8. Calibrationand key in password
13579 to enter the following calibration menu screen:
:
L