User Manual
101 
calibration menu. 
3.  First, input the resistance of the current shunt in mΩ for current 
measurement fixture and press  . 
4.  Then, observe the voltage value shown on the DVM, and input it 
accordingly for CALIB CURR Lo, then CALIB CURR MID1, 
MID2, MID3, and Hi accordingly. 
5.  Be sure to check that all values are within the range specified by 
the below tables, depending on the model. 
XLN60026     
Function 
Setting 
Range of transformation value 
Lo 
0.1352 
0.01 ~ 0.1487 
 MID1 
0.3424 
0.3082 ~ 0.3766 
 MID2 
1.04 
0.936 ~ 1.144 
 MID3 
1.7342 
1.5608 ~ 1.9076 
Hi 
2.47 
2.223 ~ 2.717 
XLN30052     
Function 
Setting 
Range of transformation value 
Lo 
0.2704 
0.02 ~ 0.2974 
 MID1 
0.6812 
0.6131 ~ 0.7493 