5. Calibration
Note: B&K Precision recommends a calibration interval of one year for
this power supply. The following calibration instructions may be used by
authorized technicians or calibration personnel only. If you are not
authorized, do not attempt to calibrate the instrument yourself, as it may
damage the instrument and void the warranty.
5.1 Required Equipment
- 5 1/2 DVM (HP34401A)
- Shunt for current calibration (100 A/ 10 mΩ)
5.2 Voltage Calibration
Follow the step by step instructions below for voltage calibration.
1. Power off the supply and connect its output terminal to a Digital
Volt Meter (DVM), as shown in the figure below. Then set the
meter for DC voltage measurement.